Sirona Dental Systems GmbH 5Important information
Service Manual INTEGO / INTEGO pro 5.12Chair movement self-test
båÖäáëÜ
5.12
Chair movement self-test
Lifting motor self-test/General
GA647
Lifting motor self-test/Technician
Explanation
If there has been no chair movement within the past 6 hours, a brief self-
test is performed when the treatment center is switched on to check the
movement of the chair.
Exceptions
The treatment center does not perform a self-test under the following
circumstances:
The treatment center is in test mode.
An incompatible lifting drive has been connected.
No dentist element has been connected (e.g. installation of chair
base).
The DAB board in the dentist element is not detected by the system.
A safety switch is active (e.g. cuspidor retracted).
IMPORTANT
Automatic chair movement during treatment center self-test
When the treatment center is switched on, the lifting drive briefly moves
the chair up and down as part of a self-test.
The lifting drive performs this self-test when the treatment center is
switched on under the following circumstances:
- When it is switched on for the first time
- If there has been no chair movement within the past 6 hours
- If no safety switch is active
IMPORTANT
Treatment center self-test not fully completed
If any safety switches are active when the treatment center is switched
on, the self-test is not completed.
Eliminate the cause, for example: Swivel the cuspidor toward the
water unit.