64 57 316 D3543
72 D3543.076.01.04.02 06.2015
6Overview of modules and boards Sirona Dental Systems GmbH
6.3Dentist element Service Manual INTEGO / INTEGO pro
6.3.10 Board DAPm - instrument control
Components on the board
LEDs
LED Color Function
36V Green 36 V input voltage
12V Green Generated +12 V voltage
5V Green Generated +5 V voltage
5V_ISO Green 5 V input voltage
3.3VANA Green Generated +3.3 V reference voltage
RUN Green Flashes 1x per second during normal operation
Error Red Flashing:
2x Communication with DSA lost
3x Output stage error or instrument error
Reset Red Lights up while the board is performing a reset
36V
12V
5V
3,3VANA
BOOT
RUN
ERROR
RESET
INSTR
INST_ACTIVE
LIGHT
FS_ACTIVE
5V_ISO
CAN+
CAN -
X4 / X9
X2
X3
36V
GND