7Replacing components Sirona Dental Systems GmbH
7.1General: Service Manual INTEGO / INTEGO pro
7
Replacing components
7.1
General:
NOTICE
Disconnecting a component from the chair
A large volume of water can escape from the connections when
disconnecting a component.
Drain the disinfectant tank (see Disinfectant tank [ 119]) and the
mixing tank (see Mixing tank [ 118]).
Before disconnecting the dentist element, assistant element, or
water unit from the chair, switch the treatment center off at the main
switch.
NOTICE
Modules that carry water and air may still be pressurized after switching
off the treatment center.
The pressure of the lines can be relieved by unscrewing the water
and air filters. The filters can be accessed via the maintenance flap
of the water unit.
IMPORTANT
When refitting the foot switch or BOF variant, resetting to factory
settings is a must!
Please Inform the user that the user-specific settings are reset to default
values by this reset.