EasyManua.ls Logo

Sirona Intego - Technical data

Sirona Intego
162 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
64 57 316 D3543
D3543.076.01.04.02 06.2015
25
Sirona Dental Systems GmbH 5Important information
Service Manual INTEGO / INTEGO pro 5.4S-COM
båÖäáëÜ
5.4
S-COM
5.4.1 General
S_COM
On the S-COM, the PC software functions are performed for the user
interface and the user interface software is downloaded via FTP.
If the S-COM connection has been established, the S-COM LED lights up
on the DSA and on the DAK. If this LED is flashing, the connection is
being established or a malfunction has occurred. However, there is a line
connection. If the LED is off, there is no data connection. In this case, it
must be assumed that there is a disruption in the line.
The board DAT has a 100 Ohm terminating resistor for the S-COM so that
the line is closed if a board DAK is not fitted.