www.spacelabshealthcare.com
1-11
91390 QUBE
I NTRODUCTION
Figure 1-4 Battery charger and AC/DC power indicators
2
1
Table 1-4 Power Indicators (only shown when AC is connected)
Description Symbol on panel
1 AC/DC Power Indicator - External power supply is connected
to the monitor and the AC mains. When the AC power
indicator comes on, the connection is good.
Refer to Figure 1-4.
2 Battery Charge Status Indicator-
Shown if a battery is present. The left light shows if the top
battery is installed. The right light shows if the bottom
battery is installed.
Flashing (rate of 1 second on and 1 second off) shows the
battery is charging.
Steady ON shows the battery is charged.
Flashing, rapid or erratic shows the battery is faulty and must
be replaced.
Refer to Figure 1-4.
Note:
The option C monitor has one battery, so only one
battery charge indicator shows.

Table of Contents