4-16
91390
QUBE
M AINTENANCE
Figure 4-9 Main Diagnostics Menu
The process takes about 10 seconds to complete. The results are
shown in Figure 4-10 on page 4-16.
Figure 4-10 Format USB Drive Results Screen
6 To cold-boot the monitor, touch or click the D - run diagnostics,
then the R - Reset Monitor (cold boot) keys. Let the monitor to
fully boot.
7 To restore the faculty user-defined rends and DNA (Citrix)
certificates, refer to the XPREZZON and qube Bedside Monitoring
Operations Manual (P/N 070-2112-xx).
8 Perform the retest procedure. Refer to Functional Tests on
page 4-5.