EasyManua.ls Logo

Stryker LUCAS 3 - View Error Code

Stryker LUCAS 3
91 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LUCAS
®
3 Chest Compression System Service Manual
3328798-002, ©2017 Physio-Control, Inc.
Page 22 of 91
View Error Code
Once the connection is made from Bluetooth or USB cable, the information about the device should appear in the
designated CPU windows.
The device Errors and Warnings will appear in each CPU’s box window.
To view existing log information on the device, click LOG button,
, Log view shows log data on screen.
Highlight a row
Press READ to view log data
Press USE to see the Error and Warnings code in different CPU windows
Press SAVE to save the raw log file data to PC (only for developers to use)
Press EXIT to exit LOG view
Example:
1. Highlight a row according to Date and Time the error occurred from LOG view;
2. Click READ and wait to see the “Collect errors and warnings from LOG File” to show up on the left side of
window;
3. If the Number of Errors or Warnings are not 0, Click USE to see the highlighted LOG Error code in the CPU
windows (USE will exit LOG view also).
- Picture shows LOG File content on the left for LUCAS 3 Version 3.0 (Part No: 250039-00)
Note, differences in J-numbers and CheckSum Numbers may occur due to different program versions
3.
2.
1.

Table of Contents

Other manuals for Stryker LUCAS 3

Related product manuals