LUCAS
®
3 Chest Compression System Service Manual
3328798-002, ©2017 Physio-Control, Inc.
Page 56 of 91
Preventive Maintenance
Maintenance Procedure
NOTE: Use PIP checklist 3326789 Rev C to record PIP results during the maintenance procedure.
Step
Routine
Description
Ref. /
1
Cleaning
Clean the device according to IFU, replace if
necessary
6.1 in the
IFU
2
Check/Replacement Bellows
Check Bellows is intact and clean, replace if
necessary according to How to Replace the Bellows
3
Check/Replacement Suction Cup
Check Suction Cup is intact and clean, replace if
necessary
6.2 in the
IFU
4
Check/Replacement Patient
Straps
All fabrics (Patient Straps, Stabilization Strap and the
Carrying Bag) are checked with the aspect of
cleanliness and that the Velcro and buckles isn’t worn
and fulfils its function. Replace if necessary.
6.3 in the
IFU
6.4 in the
IFU
5
Check/Replacement Stabilization
Strap
6
Check/Replacement Carrying
Bag
7
Check/Update Software
Check the software version in the device and update if
possible according to LUCAS® 3 Program Loader
8
Check for Errors and Warnings
Start the device and let it run for a minute to check
that no errors or warnings occur according to
Electronics Test
9
Clean Fan and Mesh Grill
According to Clean Electric Fan and Mesh Grill
10
Compression Module Lubrication
According to Compression Module Lubrication
11
Claw lock lubrication
According to Claw Lock Lubrication
12
Function Check
Perform a function check according to Function Check

Table of Contents

Other manuals for Stryker LUCAS 3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker LUCAS 3 and is the answer not in the manual?

Stryker LUCAS 3 Specifications

General IconGeneral
Device TypeChest Compression System
Power SourceRechargeable Lithium-ion Battery
Compression Depth5.3 ± 0.2 cm (2.1 ± 0.1 in)
Battery Life45 minutes of operation
Battery Charge TimeApproximately 4 hours
Operating ModesPause
Patient SizeFits most adult patients
Operating Temperature0°C to 40°C (32°F to 104°F)
Storage Temperature–20 to +60 °C (–4 to 140 °F)
IP RatingIP43
Compression Rate100 compressions per minute