LUCAS
®
3 Chest Compression System Service Manual
3328798-002, ©2017 Physio-Control, Inc.
Page 29 of 91
- LUCAS 3 Version 3.1 (Part No: 250041-00)
Calibrate Linear Sensor
After software update is complete, Click POT CAL to calibrate Linear Sensor. Before you click the Start button please
make sure that nothing can influence the movement of the suction cup, that the device is in its upright position and
that the Suction Cup is approximately 20mm (~1 inch) down from its upmost position.