LUCAS
®
3 Chest Compression System Service Manual
3328798-002, ©2017 Physio-Control, Inc.
Page 48 of 91
COM_B_TIMEOUT
No=0x55
I2C Battery
Charger
Communication
Timeout
No I2C communication
with Charger system
>100s.
I2C communication
fault or halted Charger
CPU.
Check cable
between boards.
Restart LUCAS3
and run for 30
minutes. If fault
repeats replace
Protective board.
Device
cannot be
operated.
Switch off by
removing
battery.
COM_I2C
N0=0x56
I2C faults too high
Disturbed I2C
communication.
Check cable
between boards.
Protective or
Control board.
Poor
response to
keystrokes.
Error Codes from Protective CPU
Error Code
Explanation
Probable Cause
Module to check
/ replace
User
Notification
LRG Text
T1_RAM
Internal RAM
test detect fault
at start.
Protective CPU fault.
Protective board
High priority
Alarm.
Internal fault
- contact
Technical
Support
T1_E2_PROM
Internal E2 prom
test detect fault
at start.
Protective CPU fault.
Protective board
High priority
Alarm.
Internal fault
- contact
Technical
Support
T1 PISTON
REVERSE ERROR
Reverse test not
done correct at
start
Motor not moved
correct or fault
current read.
Protective
board, check
cable between
boards or
Control board
(P45 fault)
After 3 start
attempt, High
priority
Alarm.
Self-test
fault
T1 PISTON
RELEASE ERROR
Protective
release line
operation fault at
start
Current detected at
disable test or moved
too low at enable test.
Check cable
between
boards,
Protective board
or Control board
(P45 fault)
After 3 start
attempt, High
priority
Alarm.
Self-test
fault
TOO_DEEP
COMPRESSION
Too deep
compression
detected in
active mode.
Protective
potentiometer fault.
Jam of compression
module at run with no
load.
Check
smoothness in
Compression
module,
Protective
potentiometer.
High priority
Alarm.
In Pause:
Disruption of
Suction Cup
In Active:
Compressio
ns out of
range
TOO_SHALLOW
COMPRESSION
Too shallow
compression
detected in
active mode.
(50 consecutive
strokes)
Protective
potentiometer fault.
Compression module
runs with too high load.
Check load and
Protective
Potentiometer at
different heights.
High priority
Alarm.
Compressio
ns out of
range
INTERNAL
TEMPERATURE
Protective board
temperature
detected
> 85°C
Fan or Protective
board sensor or
voltages fault
Fan, Protective
board. Check
ambient
temperature
(below +40°C).
High priority
Alarm.
Too high
temperature

Table of Contents

Other manuals for Stryker LUCAS 3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker LUCAS 3 and is the answer not in the manual?

Stryker LUCAS 3 Specifications

General IconGeneral
Device TypeChest Compression System
Power SourceRechargeable Lithium-ion Battery
Compression Depth5.3 ± 0.2 cm (2.1 ± 0.1 in)
Battery Life45 minutes of operation
Battery Charge TimeApproximately 4 hours
Operating ModesPause
Patient SizeFits most adult patients
Operating Temperature0°C to 40°C (32°F to 104°F)
Storage Temperature–20 to +60 °C (–4 to 140 °F)
IP RatingIP43
Compression Rate100 compressions per minute

Related product manuals