LUCAS
®
3 Chest Compression System – Service Manual
3328798-002, ©2017 Physio-Control, Inc.
Page 86 of 91
Replacing the Main Body/Frame Assembly
4. Disassembling
• Follow the instructions in Section Replace the Compression Module.
• Follow the instructions in Section Replace the Controller PCB
• Follow the instructions in Section Replace the Protective PCB
• Follow the instructions in Section Replacing the Power Inlet.
• If replacing the Main Body follow the instructions in Section Replace the Support Legs.
Warning: Always use ESD protection when handling PCB’s!
5. Reassembling
• Re-install the Power Inlet using instructions in section Replacing the Power Inlet.
• Install the Main Body Lid and tighten all four Torx 20 screws to 13 in-lbs / 1.5 Nm.
• Re-install the Compression Module, Controller PCB, Protective PCB, and Hood.
• If removed, re-install the Support Legs.
6. Test
• Perform a Function Check according to Section Function Check.