LUCAS
®
3 Chest Compression System Service Manual
3328798-002, ©2017 Physio-Control, Inc.
Page 86 of 91
Replacing the Main Body/Frame Assembly
4. Disassembling
Follow the instructions in Section Replace the Compression Module.
Follow the instructions in Section Replace the Controller PCB
Follow the instructions in Section Replace the Protective PCB
Follow the instructions in Section Replacing the Power Inlet.
If replacing the Main Body follow the instructions in Section Replace the Support Legs.
Warning: Always use ESD protection when handling PCB’s!
5. Reassembling
Re-install the Power Inlet using instructions in section Replacing the Power Inlet.
Install the Main Body Lid and tighten all four Torx 20 screws to 13 in-lbs / 1.5 Nm.
Re-install the Compression Module, Controller PCB, Protective PCB, and Hood.
If removed, re-install the Support Legs.
6. Test
Perform a Function Check according to Section Function Check.