Failed calibration
Case 1:
If the value for Control exceeds 6% or if the value for Protective exceeds 3.1%, the device will make an alarm and an
error code will be shown in the Control CPU’s window, the sensor calibration has failed.
To recalibrate:
1. Mute the alarm
2. Check that nothing interferes with the movement of the suction cup
3. The device is upright
4. Click POTCAL to make a new calibration.
If the second calibration also fails it might be necessary to replace the Compression Module.
Case 2:
If this error “Cup not moved to top” showed in Control CPU window, manually move the suction cup down an inch
and then click POT CAL again to recalibrate.