LUCAS
®
3 Chest Compression System Service Manual
3328798-002, ©2017 Physio-Control, Inc.
Page 49 of 91
PISTON_TIME_OUT
Protective
potentiometer
not detected a
movement of
50% stroke for
10s.
Protective
potentiometer fault.
Too high load. I2C
communication fault at
mode change.
Check load and
Protective
potentiometer at
different heights.
Check for
communication
timeouts.
High priority
Alarm.
Disruption of
Suction Cup
CONTROL_LIFETICK
I2C
communication
fault with Control
CPU
(Warning)
I2C communication
fault at mode change.
Other fault detected by
Control.
Check if other
fault is
detected by
Control. Check
cable between
boards.
None.
Warning
stored to
LOG file.
ALARM TYPE ROM
TEST
Internal ROM
test detect fault
at start.
Protective CPU fault.
Protective board
High priority
Alarm.
Internal fault
- contact
Technical
Support
CHARGER_STOP
Charger fault
detected
(Motor disable)
See Charger errors
See Charger
errors
High priority
Alarm.
CONTROL STOP
Control fault
detected
(Motor disable)
See Control errors
See Control
errors
High priority
Alarm.
Error Codes from Charger CPU
Error Code
Explanation
Probable Cause
Module to check
/ replace
User Notification
LRG Text
T1 RAM
ALARM
Internal RAM
test detect fault
at start.
Charger CPU
faults.
Protective board
High priority Alarm.
Internal fault -
contact
Technical
Support
T1 E2 PROM
ALARM
Internal E2 prom
test detect fault
at start.
Charger CPU
faults.
Protective board
High priority Alarm.
Internal fault -
contact
Technical
Support
T1 POWER
OFF ALARM
T1 test not
ended correct.
T1 fault detected by
other CPU’s
Check other
faults.
After 3 start attempt,
High priority Alarm.
Self-test fault
BATT HIGH
TEMP
ALARM
Battery
temperature
detected to be
>70°C
Battery fault.
Ambient
temperature too
high.
Battery
High priority Alarm.
Battery too high
temperature
(S/N xxxxxxxx)
BATT
EMPTY
ALARM
Battery
continues below
25V
Uncharged battery.
Battery, DCIN or
charging fault.
Battery, Protective
board, DCIN
adaptor
High priority Alarm.
Battery
depleted
(S/N xxxxxxxx)
BATT 10 MIN
ALARM
Battery
continues below
25.3V
Uncharged battery.
Battery
Medium priority
Alarm. Flashing
orange battery LED
(B1)
Battery low
charge
(S/N xxxxxxxx)
T1 ROM
ALARM
Internal ROM
test detect fault
at start.
Charger CPU faults.
Protective board
High priority Alarm.
Internal fault -
contact
Technical
Support
T1 BATT
COM
ALARM
Battery
communication
fault
Connector fault.
Check
connections.
Unplug and insert
battery, check for
fault. Or replace
battery.
None.
Shall be shown at
LOG file read.
Battery not
recognized as
Physio-Control
proprietary (no
S/N)

Table of Contents

Other manuals for Stryker LUCAS 3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker LUCAS 3 and is the answer not in the manual?

Stryker LUCAS 3 Specifications

General IconGeneral
Device TypeChest Compression System
Power SourceRechargeable Lithium-ion Battery
Compression Depth5.3 ± 0.2 cm (2.1 ± 0.1 in)
Battery Life45 minutes of operation
Battery Charge TimeApproximately 4 hours
Operating ModesPause
Patient SizeFits most adult patients
Operating Temperature0°C to 40°C (32°F to 104°F)
Storage Temperature–20 to +60 °C (–4 to 140 °F)
IP RatingIP43
Compression Rate100 compressions per minute