LUCAS
®
3 Chest Compression System Service Manual
3328798-002, ©2017 Physio-Control, Inc.
Page 83 of 91
Remove the Communication PCB
2. Reassembling
Put the new Communication PCB in place with the new plastic screws, plastic spacers, and the nuts
Finger tightens screws by holding screw head lightly with fingertip until all items are tight against each
other and no play is visible. Do not overtighten.
Then use tools to tighten screw and nut ¾ of turn (270), tightening order according to the numbers shown
in the picture
Flatten screws (3x) using a non-knurled flat nose plier right up against the nut shown in the picture below
Put back the Controller PCB according to Reassembling instruction in
How to Replace the Controller
PCB
3. Programming
Connect a laptop and program the Protective PCB according to LUCAS® 3 Program Loader
.
Check that “Device S/N”, Time and Date, “Total Number of Strokes”, “Hardware ID” and “Device Friendly
Name” are correct (if possible check against “Store View” performed in Disassembling step). If “Store View”
was not possible set: “Device S/N” (e.g. 35160021) and “Hardware ID” (e.g. 100921-00) according to Type
Label, “Total Number of Strokes” to 0 (zero) and “Device Friendly Name” to “LUCAS3- + serial number
according to Type Label.
4. Test
Perform a Function Check according to Function Check
.