LUCAS
®
3 Chest Compression System Service Manual
3328798-002, ©2017 Physio-Control, Inc.
Page 46 of 91
T1_C_LOWCURR
No=0x24
Too low Current
detected at first
Current test.
Motor wires loose.
HALL sensor signal
loose or delayed
internal voltages.
Motor, HALL
cable. Check for
internal voltages
warnings. Control
board
Prolonged
startup test
time.
T1_P_DISABLE
No=0x25
Protective Disable
at current test at
first start.
P45 delayed or loose
cable between boards.
Cable between
boards.
Protective or
Control board.
Prolonged
startup test
time.
T1_P_POT
No=0x27
Read Protective
POT time out at
first start.
I2C communication
delayed. Halted
Protective CPU at first
start.
Cable between
boards. Check
Protective
warnings.
Prolonged
startup test
time.
T1_C_POT
No=0x28
Miss match at
warm start.
Control and Protective
POT value divergence
> 10mm at warm start.
POT fault or un- linear.
Preform a new
POT calibration.
Check for
voltages faults.
LUCAS V2
operation.
T1_P_STATE_5
No=0x29
No end of
Protective Current
Test (time out) at
first start.
Protective current read
fault at first start.
I2C communication
interference.
Check Protective
voltages.
Protective board.
Prolonged
startup test
time.
T1_B_OFF
No=0x2A
Battery Charger
not answered T1
test done (time
out) at first start.
Charger CPU not done
T1 test correct or I2C
communication faults.
Check Charger
voltages.
Protective board.
Prolonged
startup test
time.
T1_C_DOWN
No=0x2D
Control not
moving down
before REW test
at first start.
Motor connections.
Protective stop by
disable line. High load
at start test.
Motor. Check if
start test is done
with load.
Prolonged
startup test
time.
T1_P_REW
No=0x2E
Protective Time
out at Reverse
test at first start.
Motor connections.
Protective POT signal.
Locked compression
module.
Motor. Protective
POT. Use free
piston movement
at start test.
Prolonged
startup test
time.
T1_C_TOTOP
No=0x2F
Control Time out
at move to top at
first start.
Motor and HALL
connections.
Locked compression
module.
Motor.
Compression
module. Use free
piston movement
at start test.
Prolonged
startup test
time.
INT_P_5V
No= 0x30
Protective 5V <
4.5V or > 5.5V
Cable between
boards. Incorrect
Protective 5V. Internal
reference faults.
Check cable.
Protective 5V.
Control internal
voltages faults.
Protective board.
None.
INT_C_5V
No=0x31
Control 5V < 4.5V
or > 5.5V
Electrical fault. Control
board pin connected to
back plate.
Control board.
Check for other
Control warnings.
None.
INT_C_24V
No=0x32
Control 24V < 18V
Battery fault. Internal
reference faults.
Check other
Control internal
voltages faults.
Test with a new
charged battery.
None.
LOG file
stopped.
INT_C_REF
No=0x33
Control 2.5V Ref <
2V or > 3V
Electrical fault. Control
board pin connected to
back plate or POT
+2.5V shortcut.
Check reference
voltage or
replace Control
board.
None.
INT_C_TEMP
No=0x34
High internal
board temp.
> 70°C Warning
Fan disconnected or
jammed. Broken Temp
sensor or internal
reference faults.
Fan. Control
PCBA
Temperature
warning
signal.
High
temperature

Table of Contents

Other manuals for Stryker LUCAS 3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker LUCAS 3 and is the answer not in the manual?

Stryker LUCAS 3 Specifications

General IconGeneral
Device TypeChest Compression System
Power SourceRechargeable Lithium-ion Battery
Compression Depth5.3 ± 0.2 cm (2.1 ± 0.1 in)
Battery Life45 minutes of operation
Battery Charge TimeApproximately 4 hours
Operating ModesPause
Patient SizeFits most adult patients
Operating Temperature0°C to 40°C (32°F to 104°F)
Storage Temperature–20 to +60 °C (–4 to 140 °F)
IP RatingIP43
Compression Rate100 compressions per minute