2. Reassembling
Apply Thread Lock Fluid on two of the screws and place them in the bracket according to picture below
Apply Thread Lock Fluid to the other two screws and put them on the Motor together with the lock washer
and terminal. Make sure that the cable outlet is oriented as figure below.
Slide the Motor onto the Bracket and tighten the screws to 18.8 in-lbs / 2.0 Nm with a Torque Screwdriver
Torx 20 (T20), and make sure to angle the terminal so the ferrite of the Hood Power Cable is secured
behind the Hood Ground Cable connector.