Chapter 5 Command Reference
Description of SCPI Commands
Operating and Programming Manual 79
DIAGnostic:CURRent :BEAM?
This query returns the beam current setpoint in amperes.
Response format = <NR3>, precision = 1 x 10
-9
A.
DIAGnostic:CURRent:CFIeld?
This query returns C-field current-setpoint value in amperes.
Response format = <NR3>, precision = 1 x 10
-6
A.
DIAGnostic:CURRent :PUMP?
This query returns the ion pump current in amperes.
Response format = <NR3>, precision = 2 x 10
-7
A.
DIAGnostic:GAIN?
This query returns counter signal gain as fraction of full-scale value.
Response format = <NR3>, precision = 1 x 10
-3
.
DIAGnostic:LOG:CLEar
This command purges all event log entries.
The 5071A must be in REMote mode to execute this command. This command is
not affected by *RST and does not have a query.
DIAGnostic:LOG :COUNt?
This query returns the number of entries in the log as an integer. (If the log has
been deleted, this query returns a negative value.)
Response format = <NR1>.
DIAGnostic:LOG :PRINt?
This query prints the entire contents of the event log. (Equivalent to the front
panel ―LOG PRINT‖ command.)
Response format = <arbitrary ASCII>
DIAGnostic:LOG[:READ]? [<numeric_value>]
This query returns the time stamp and text of the most recent unread log entry if
no parameter is supplied; otherwise, it returns the specified log entry.
Response format = <string quoted, string quoted>.