Chapter 5 Command Reference
Description of SCPI Commands
80 Operating and Programming Manual
DIAGnostic:LOG:VERBosity DISable|TERSe|VERBose|SERVice
This command specifies the log stream events actually sent to the serial port.
The 5071A must be in REMote mode to execute this command.
The following four parameters of the DIAGnostic:LOG:VERBosity command
allow you to select the log stream events that will be sent to the serial port:
• DISable prevents messages from being echoed to the serial port.
• TERSe causes only fatal errors to be echoed to the serial port.
• VERBose echoes all progress information and user events.
• SERVice adds detailed power-on/warm-up events to VERBose content.
The VERBosity setting has no effect on which messages are saved in non-
volatile log. It only specifies which messages echo to the serial port unrequested.
DIAGnostic:LOG:VERBosity?
This query reports the active setting for log verbosity.
Return values = DIS, TERS, VERB, or SERV, (discussed above).
DIAGnostic:RFAMplitude?
This query returns RF Attenuator setpoints as fraction of full-scale.
Response format = <NR3, NR3>, precision = 1 x 10
-3
.
DIAGnostic:STATus[:GLOBal]?
This query returns instrument‘s summary status message.
Response format = <string quoted>.
DIAGnostic:STATus:SUPPly?
Note: The response ―BATT‖ does not apply to 48V option.
This query returns a power supply status indication.
Response format = AC|DC|BATT|LOW, defined as follows:
• AC - operating from ac power.
• DC - operating from auxiliary dc power.
• BATT - operating from internal battery backup.
• LOW - internal supply voltages below nominal.