Chapter 5 Command Reference
Description of SCPI Commands
86 Operating and Programming Manual
PTIMe:LEAPsecond[:STATe] OFF|0|ON|1
This command activates or cancels a programmed leap second.
When STATe is turned ON, a leap second is scheduled for the last specified
values of DURation and MJDate.
The function of this command is not affected by *RST.
PTIMe:LEAPsecond[:STATe]?
This query indicates whether a leap second is pending.
Response format = <0 or 1>
Returns 1 if pending, 0 if none pending.
PTIMe:SLEW <numeric_value>
This command adjusts the epoch of the clock with the supplied parameter.
The 5071A must be in REMote mode to execute this command.
One parameter is required, accepts ―S‖ suffix.
Parameter range = -0.5 S to 0.5 S., precision = 50 ns.
This is an event.
The function of this command is not affected by *RST.
PTIMe:SLEW? MIN|MAX
This query returns the smallest and largest slew value which can be used.
Response format = <NR3>.
PTIMe:STANdby OFF|0|ON|1
This command enables or disables STANDBY mode.
The 5071A must be in REMote mode to execute this command.
Disabling standby initiates normal operation, if possible. (This command is
equivalent to the 5061B ―cesium off‖ switch.)
The function of this command is not affected by *RST.