Chapter 5 Command Reference
Description of SCPI Commands
Operating and Programming Manual 87
PTIMe:STANdby?
This query returns 1 if instrument is in standby, 0 otherwise.
Response format <0 or 1>
Returns 1 if in standby, 0 if not in standby.
PTIMe:SYNChronization OFF|FRONt|REAR
This command arms the specified sync port with the supplied parameter.
The 5071A must be in REMote mode to execute this command.
Arming terminates when:
• canceled from the command interface,
• a valid sync pulse is received at the armed port, or
• after 1-1/2 seconds has elapsed.
Examine the relative timing of the pulses to determine which event has occurred.
The function of this command is OFF at *RST.
The available parameters are:
• OFF — Turns off either Front or Rear Sync Input port connectors.
• FRONt — enables the Front-panel Sync Input port connector.
• REAR — enables the Rear-panel Sync Input port connector.
PTIMe:SYNChronization?
This query indicates which sync port is armed, if any.
Response format = OFF|FRON|REAR.