2. General Description
2008-07 Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4 31
Washing (i-Control):
1. Adjust the parameters on the Wash tab of the Injector Maintenance
dialog box in the Settings menu
2. Start the washing procedure by clicking the Start wash.
Wash
Select one of the injectors Injector A or Injector B or
Injector A and B.
Select the Piston Strokes (1 – 60; 1 stroke equals 1 ml)
Select the Wash Speed (100 - 300 µl/sec).
Select the Refill Speed (100 – 300 µl/sec.) or select Refill
Speed equal to Wash Speed.
Click Start wash to start the wash procedure.
Click Close to exit the dialog box.