5. Optical System
2008-07 Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4 85
Example 2:
Combination of microplate and cuvette measurement:
For some applications it might be necessary to compare data measured on a
microplate with cuvette data. The following example shows how to set up this
measurement in general:
Cuvette
Necessary for cuvette measurements.
Absorbance strip
(cuvette)
Up to 4 absorbance fixed wavelength strips are allowed.
Reference wavelength is only selectable when using one
absorbance fixed wavelength strip. ‘Apply blanking’ is
disabled when a reference wavelength is selected.
Select the appropriate measurement parameters
(wavelength, number of flashes and settle time)
Plate
Necessary for microplate measurements. Select an
appropriate plate type for the measurement.
Part of Plate
(not shown)
Optional. Use the ‘part of plate’ strip if only a part of the
plate shall be measured.
Absorbance strip
(microplate)
Up to 10 absorbance fixed wavelength strips are allowed.
Reference wavelength is only allowed on the first
absorbance strip. Reference wavelength is disabled on
absorbance strips 2 to 10.
Select the appropriate measurement parameters
(wavelength, number of flashes and settle time) for your
application.