Options and
Accessories-2230
Operators
Commands
ALL
ATRigger? [MODe]
ALL
SGLswp ARM
ALL
SGLswp?
ALL
TRlggerd?
7-28
Table 7-24
Trigger Commands
Description
Returns the present A trigger mode
in
the form: ATR MOD:<NOR,
PPA,
or
SGL>;.
PPA
is
returned for both Peak-to-Peak Auto
and
TV
Field
trigger
modes. The reply
is
the same with or without the optional [MODe] argument.
Rearms a completed single sweep.
An
execution error
is
generated if the
instrument is not
in
SGL SWP mode,
and
an
execution warning
is
generated if
the single sweep
is
already armed. With
OPC
ON,
a service request status byte
for operation complete
is
generated when the single sweep occurs.
Returns the state of the SGL SWP trigger mode
as:
SGL
<ARM
or
DON>;
when SGL SWP trigger mode
is
on.
If SGL SWP trigger mode
is
not on, a reply
of "SGL ;"
is
made,
and
an
execution warning
is
generated.
Returns the present state of the TRIG'D indicator
as:
TRI
<ON
or
OFF>;.
--

Table of Contents

Other manuals for Tektronix 2230

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2230 and is the answer not in the manual?

Tektronix 2230 Specifications

General IconGeneral
BrandTektronix
Model2230
CategoryTest Equipment
LanguageEnglish

Related product manuals