EasyManuals Logo
Home>Tektronix>Test Equipment>2230

Tektronix 2230 Operator's Manual

Tektronix 2230
188 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #86 background imageLoading...
Page #86 background image
Basic Applications-2230 Operators
1.
Preset instrument controls
and
obtain a baseline
trace.
2.
Apply a 5-division signal to either vertical-channel
input connector
and
set the VERTICAL MODE switch to
display the channel used. Ensure that the VOL TS/DIV
Variable control
is
in
the CAL detent.
NOTE
For rise times less than 0.2 µs
per
division, the
VOL
TS/DIV Variable control may be used to obtain
an exact 5-division display.
3. Vertically position the trace so that the zero refer-
ence of the waveform touches the 0% graticule line
and
the top of the waveform touches the 100% graticule line
(see Figure 6-28).
ULY
=1!:>
•
l!Jr
is
i
I
1~0
j
-
I
,0
V
T
V
i\.A
V
~
/
'°
I
-'!
I
r
l?OIM'~
'i0ns
THE
10%
POINT ON THE WAVEFORM INTERSECTS
THE CENTER VERTICAL GRATICULE LINE.
THE
90%
POINT ON THE WAVEFORM INTERSECTS
THE CENTER VERTICAL GRATICULE LINE.
4998-41
Figure 6-28. Rise time, differential time method.
4.
Set the A SEC/DIV switch for a single-waveform
display. Ensure that the A
and
B SEC/DIV Variable control
is
in
the CAL detent.
6-28
5.
Select BOTH HORIZONTAL MODE
and
set the B
SEC/DIV switch
to
spread the rise-time-measurement
portion of the display
as
much
as
possible.
6.
Select the B HORIZONTAL MODE. Adjust the B
DELAY TIME POSITION control until the display intersects
the 10% point at the center vertical graticule line (see Fig-
ure 6-28, Point
A).
7.
Record the DELAY TIME POSITION readout.
8.
Adjust the B DELAY TIME POSITION control until
the display intersects the 90% point at the center vertical
graticule line (see Figure 6-28, Point
B).
9.
Record the DELAY TIME POSITION readout.
10. Calculate rise time using the same formula listed
in
the "Time Difference Between Repetitive Pulses" measure-
ment procedure.
Time Difference Between Two Time-Related
Pulses
1.
Preset instrument controls
and
obtain a baseline
trace.
2.
Using probes or cables having equal time delays,
apply the reference signal to the Channel 1 input
and
apply the comparison signal to the Channel 2 input.
3.
Set both VOL TS/DIV switches to produce a display
of 2 to 3 divisions
in
amplitude.
4.
Select BOTH VERTICAL MODE
and
either ALT or
CHOP, depending on the frequency of the input signals.
5.
Set the A SEC/DIV switch to display the measure-
ment points of interest within the graticule area.
6.
Select
BOTH
HORIZONTAL MODE
and
CH
1
VERTICAL MODE. Adjust both the Channel 1 POSITION
control
and
the A/B SWP
SEP
control so that the A trace
is displayed above the B trace.
--

Table of Contents

Other manuals for Tektronix 2230

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2230 and is the answer not in the manual?

Tektronix 2230 Specifications

General IconGeneral
BrandTektronix
Model2230
CategoryTest Equipment
LanguageEnglish

Related product manuals