Pinpoint Trigge
rs
Trigger on B Event
The A trigger arms the instrument.
Posttrigger acquisition starts on the nth B
event.
B Trigger After Delay Time
The A trigger arms the instrument.
Posttrigger acquisition starts on the first B
edge after the trigger delay time.
Triggering with Reset
You can specify a condition to reset the
trigger system if it occurs before the B trigger
event. When the reset event occurs, the
trigger system stops waiting for the B event
and returns to waiting for the A event.
Quick Tips
B-trigger delay time and horizontal delay time are independent functions. When you establish a trigger condition using
either the A trigger alone or the A and B triggers together, you can also use horizontal delay to delay the acquisition
by an additional amount.
34 TDS6000B & TDS6000C Series Quick Start User Manual