6-11
Troubleshooting
H/S PROBEH/S PROBE
MessageMessage Meaning and Corrective ActionMeaning and Corrective Action
Connect (Arterial or
Venous) H/S to Cuvette
H152
The H/S Probe detects when it is still connected to
a color chip in Measurement Mode. The system will
not report values while connected to a color chip.
Remove the H/S Probe from the color chip and
attach the Cuvette to obtain measurement values.
Technical Alarm
Cuvette not detected on
(Arterial or Venous) H/S
The system will alarm when measurement is active
and a Cuvette is not detected by the H/S Probe.
Connect the H/S Probe to the Cuvette. Ensure the
release lever of the H/S Probe is latched by pressing
the Probe and Cuvette together. If the alarm
continues, replace the Cuvette and contact Terumo
Cardiovascular Technical Support.
Note: Connecting the H/S Probe to the Cuvette will
deactivate the alarm.
FLOW INTERFACE MODULE AND SENSORFLOW INTERFACE MODULE AND SENSOR
MessageMessage Meaning and Corrective ActionMeaning and Corrective Action
(Name)
Flow Module-[SN] error
F001
F004
F006
F007
F008
F010
F011
F012
F013
F014
F055
F204
F205
F206
F207
F208
A problem was detected in the Flow Interface
Module electronics.
It may be possible to reset the Flow Interface
Module by disconnecting and reconnecting it to
the Core. If the issue persists, use a replacement
Flow Interface Module and contact Terumo
Cardiovascular Technical Support.