6-12
Troubleshooting
FLOW INTERFACE MODULE AND SENSORFLOW INTERFACE MODULE AND SENSOR
MessageMessage Meaning and Corrective ActionMeaning and Corrective Action
Flow Sensor-[SN]
measurement error
F015
This error indicates invalid Flow Sensor
measurement readings.
Try a replacement Flow Sensor. If the issue persists,
contact Terumo Cardiovascular Technical Support.
Check Flow Interface
Module-[SN] connection
F050
F051
F052
F053
Data from the Flow Sensor is not being received or
is invalid.
Check the connection between the Interface Module
and Flow Sensor, or use a replacement Flow Sensor
and/or Interface Module. If the issue persists,
contact Terumo Cardiovascular Technical Support.
Flow Sensor-[SN] error
F200
A problem was detected in the Flow Sensor
electronics.
Try disconnecting and reconnecting the Flow Sensor
to the Flow Interface Module, or use a replacement
Flow Module. Contact Terumo Cardiovascular
Technical Support.
Technical Alarm
Verify Flow Sensor-[SN]
coupling
F201
The system will alarm when in Measurement Mode
and the Flow Sensor is not properly coupled to
tubing.
Conrm the Flow Sensor is properly attached to the
tubing. Conrm the Flow Sensor size matches the
tubing (ID, OD) being used. If the alarm continues,
replace the Flow Sensor and contact Terumo
Cardiovascular Technical Support.
Note: Proper coupling of the Flow Sensor to tubing
will deactivate the alarm.
Flow Sensor-[SN]
disconnected
F202
The system detects that the Flow Sensor has been
disconnected. To measure ow, connect a Flow
Sensor.
If this message appears when the Flow Sensor is
connected to the module, try disconnecting and
reconnecting the Flow Sensor to the Flow Interface
Module, or use a replacement Flow Sensor. Contact
Terumo Cardiovascular Technical Support.