6-14
Troubleshooting
rSOrSO
22
INTERFACE MODULE INTERFACE MODULE
MessageMessage Meaning and Corrective ActionMeaning and Corrective Action
rSO
2
Interface
Module-[SN] failure
X001
X004
X006
X007
X008
X010
X011
X012
X013
X014
X055
A problem detected in the rSO
2
Interface Module
electronics.
It may be possible to reset the rSO
2
Interface
Module by disconnecting and reconnecting to
the Core. If the issue persists, use a replacement
rSO
2
Interface Module and contact Terumo
Cardiovascular Technical Support.
Check rSO
2
Interface
Module-[SN] connection
X050
X051
X052
X053
X251
Data is invalid or is not being received from
the source cerebral oximetry device while in
Measurement Mode.
Conrm the correct oximeter is selected in the case
prole. Check the connection between the Interface
Module and oximeter. Check the type of serial cable
being used; the cable must be a straight through
serial cable. Ensure the oximeter is congured per
the instructions in Chapter 3, “Sensor Connections.
If the issue persists, try a replacement Interface
Module and contact Terumo Cardiovascular
Technical Support.
DMS INTERFACE MODULEDMS INTERFACE MODULE
MessageMessage Meaning and Corrective ActionMeaning and Corrective Action
DMS Interface
Module-[SN] failure
M001
M004
M006
M007
M008
M010
M011
M012
M013
M014
M055
A problem was detected in the DMS Interface
Module electronics.
It may be possible to reset the DMS Interface
Module by disconnecting and reconnecting it to
the Core. If the issue persists, use a replacement
DMS Interface Module and contact Terumo
Cardiovascular Technical Support.