2-7
Initial Assembly and Admin Setup
Caution
Use only CDI Gas 1 and CDI Gas 2 calibration gas bottles intended
specically for use with the CDI Model 740 Calibrator. Use of any
other gases will result in calibration failure or yield inaccurate
calibration results, causing inaccurate system performance.
Do not use the gas bottles after the recommended expiration
date. Use of the gas bottles after that date may result in inaccurate
calibration values.
Contact with combustible material may cause re.
Note: Terumo Cardiovascular recommends that the gas bottles remain in the Calibrator
at all times to avoid introducing debris into the calibrator. If the Calibrator is to be
stored for a long period of time, remove the gas bottles but ensure the openings are
kept covered.
Note: The Calibrator must be placed in an upright position to ensure correct calibration.
Note: CDI gases are contained in disposable cylinders. These non-returnable cylinders
contain nontoxic, non-ammable gases and gas mixtures.
To dispose of a gas bottle:
1. To release residual pressure of the gas bottle, press inside the centermost
point of the threaded portion of the valve. For gases or mixtures containing
less than 21% oxygen, this should be done in a well-ventilated area to avoid
asphyxiation by displacement of oxygen.
2. Remove or obliterate markings that indicate the bottle contains hazardous
material.
3. Discard the bottle as you would other metallic or hard goods trash as permitted
by local authorities, rules or regulations.
Note: To dispose by recycling, the valve must also be removed from the bottle, as it is
made from a different type of metal.