Running a Case
Temperature pH/Temperature pH/αα-Stat Selection-Stat Selection
Tap the pH-stat/37° α-stat button to display arterial and venous blood gas values either
at pH-stat (actual temperature/measured) or corrected to 37°C alpha-stat. Toggle the
pH-stat/37° α-stat to change the patient temperature mode.
Note: The temperature measured by a CDI Shunt Sensor will be that of the shunt / purge
line, which may differ slightly from the temperature in the rest of the circuit because of
the exposure of the shunt line to room air temperatures.
Caution
Shunt Sensor temperatures do not reect a patient’s temperature
and should not be used for patient temperature management.
pH-Stat 37° α-stat