7 Parameterizing, adjusting and analyzing software (P2A software)
130
Field Explanation
Operating hour
probe
Operating hour of the probe at which the
change at the instrument was performed.
Unit Unit during the adjustment.
Offset from 1-point adjustment: Value before the
adjustment.
Offset to 1-point adjustment: Value after the
adjustment.
Target value 2-point adjustment: Reference value.
Reading 2-point adjustment: Value before the
adjustment.
Difference 2-
point adjustment: Difference determined by
instrument between target value and reading.
Error/status
messages
Error/status messages: The error/status
messages assigned to the probe.
The table is shown only for error and status messages that were
generated in the transmitter and were transferred and saved there
via the connection to the P2A software.
Operating hours Operating hour at which the message
appeared in the instrument.
Serial number Serial number of the instrument to which the
message is assigned