6 Commissioning
70
6.10.3.2 Wall mounting of testo Saveris 2 H2 WLAN data
loggers
The wall bracket ensures that the WLAN data logger is kept securely in place.
Mounting materials are not included in the scope of delivery. Select appropriate
mounting materials according to the required mounting location.
(Suitable for: 0572 2035 01)
1
Wall bracket with slots for
mounting materials (screws or
cable ties)
2
Securing pin
3
Padlock
4
Storage compartment for USB
port sealing plug
5
Storage compartment for probe
socket sealing plug (left/right)
6
USB cable holder, folding: push
down on the spot marked with a
circle using a screwdriver and
slide cable holder downwards.
6.10.3.3 BT Gateway for testo 182 wall mounting
Mounting instructions
The gateway must be positioned in such a way that the ambient conditions
do not interfere with the radio signal (WLAN and Bluetooth
®
). Example:
mounting in a highly metallic environment.
For integration into a local WLAN, a minimum connection strength of 75 db is
recommended.
Live cables can lead to an impairment of the radio signal. The network
infrastructure must ensure sufficient compliance with the required wireless
standards.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TESTO Saveris Series and is the answer not in the manual?

TESTO Saveris Series Specifications

General IconGeneral
Measurement ParametersTemperature, Humidity, CO2, Pressure, Differential Pressure
Humidity Range0 to 100 %RH
CO2 Range0 to 5000 ppm
CommunicationEthernet, WiFi, USB
Data StorageInternal memory
AlarmVisual and audible alarms
Power SupplyBattery, Mains power
SoftwareTesto Saveris software

Related product manuals