2
CAUTIONS FOR USE
Important cautions
Use this instrument carefully on the following patients.
• Patients who have epidemic corneitis, conjunctivitis or any other infectious disease
• Patients who are taking medications that cause light hypersensitivity.
Basic cautions
Be careful not to let the patient touch this instrument. The patient's hand may be pinched by the
movable part.
Do not touch the focusing knob while in auto focus mode. You may be injured.
When operating the chinrest up/down switch, be careful not to pinch the patient's hand. The
patient may be injured.
Electric shock may cause burns or a possible fire. Turn the power switch OFF and unplug the
power cord before replacing the fuses.
Replace only with fuses of the correct rating.
To avoid burns, do not replace the lamp with a new one immediately after it goes off because it
is still very hot and can cause burns.
To avoid injury caused by electric shock, do not open the cover. Ask your dealer for service.
To avoid electric shock, turn off the power switch and unplug the power cord and then replace
the lamp with a rated one.
Disposal
When disposing of TRC-NW8F parts, follow the local regulations for disposal and recycling.
ENVIRONMENTAL CONDITIONS FOR USE
Temperature : 10°C - 40°C
Humidity : 30% - 90% (without dew condensation)
Air pressure : 700hPa - 1060hPa
STORAGE, USAGE PERIOD
1. Environmental conditions (without package)
*Temperature : 10°C - 40°C
Humidity : 10% - 95% (without dew condensation)
Air pressure : 700hPa - 1060hPa
* THIS INSTRUMENT DOES NOT MEET THE TEMPERATURE REQUIREMENTS OF
ISO 15004-1 FOR STORAGE. DO NOT STORE THIS INSTRUMENT IN CONDITIONS
WHERE THE TEMPERATURE MAY RISE ABOVE 40°C OR FALL BELOW 10°C.
2. When storing the instrument, ensure that the following conditions are met:
(1) The instrument must not be splashed with water.
(2) Store the instrument away from environments where air pressure, temperature, humid-
ity, ventilation, sunlight, dust, salty/sulfurous air, etc. could cause damage.
(3) Do not store or transport the instrument on a slanted or uneven surface or in an area
where it is subject to vibrations or instability.
(4) Do not store the instrument where chemicals are stored or gas is generated.
3. Normal life span of the instrument:
8 years from delivery providing regular maintenance is performed [TOPCON data]