SPECIFICATIONS & PERFORMANCE
ELECTROMAGNETIC COMPATIBILITY
This product conforms to the EMC Standard(IEC 60601-1-2 Ed.3.0: 2007).
a) MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and
needs to be installed and put into service according to the EMC information provided in the
ACCOMPANYING DOCUMENTS.
b) Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL
EQUIPMENT.
c) The use of ACCESSORIES, transducers and cables other than those specified, with the
exception of transducers and cables sold by the manufacturer of the EQUIPMENT or SYS-
TEM as replacement parts for internal components, may result in increased EMISSIONS or
decreased IMMUNITY of the EQUIPMENT or SYSTEM.
d) The EQUIPMENT or SYSTEM should not be used adjacent to or stacked with other equip-
ment. IF adjacent or stacked use is necessary, the EQUIPMENT or SYSTEM should be
observed to verify normal operation in the configuration in which it will be used.
e) The use of the ACCESSORY, transducer or cable with EQUIPMENT and SYSTEMS other
than those specified may result in increased EMISSION or decreased IMMUNITY of the
EQUIPMENT or SYSTEM.
Item Parts code
Cable
shield
Ferrite
core
Length
(m)
AC power cord 44804 7002 Not used Not used 1.5
USB cable ---- Used Used 2.0
USB cable ---- Used Used 2.0
USB cable (for USB keyboard) ---- Used Not used 2.0
USB cable (for USB mouse) ---- Used Not used 1.5
PHO.CAM_CNT D90 (for remote control) 40508 69700 Not used Not used 0.35
PHO.CAM_CNT D90 (for syncro) 40508 69700 Used Not used 0.35
PHO.CAM_POW_CBL 40508 66300 Not used Used 0.35
PHO.CAM_USB_PC 40508 69200 Used Used 1.0
Guidance and manufacturer's declaration - electromagnetic emissions
The TRC-NW8F/TRC-NW8F plus is intended for use in the electromagnetic environment speci-
fied below.
The customer or the user of the TRC-NW8F/TRC-NW8F plus should assure that it is used in
such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1 The TRC-NW8F/TRC-NW8F plus uses RF energy
only for its internal function. Therefore, its RF emis-
sions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B The TRC-NW8F/TRC-NW8F plus is suitable for use
in all establishments including domestic and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC61000-3-3
Complies