EasyManua.ls Logo

Topcon TRC-NW8F - Patients Environment; Electric Rating

Topcon TRC-NW8F
104 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
84
SPECIFICATIONS & PERFORMANCE
PATIENT’S ENVIRONMENT
When the patient or inspector may touch the devices (including the connecting devices) or
when the patient or inspector may touch the person that comes into contact with the devices
(including the connecting devices), the patient's environment is shown below.
In the patient's environment, use the device conforming to IEC60601-1. If you are compelled to
use any device not conforming to IEC60601-1, use an insulation transformer or the common
protective earth system.
ELECTRIC RATING
Source voltage : AC100-240V
Frequency : 50-60Hz
Power input : NORMAL 100VA MAXIMUM 400VA
Do not use the power strip in the
patient's environment. Connect the
power supply of the device to the
commercial power supply.
Radius 1.5m
1.5m
2.5m
1.5m
Devices applicable to the
patient's environment
Digital single-lens reflex cam-
era
USB memory
Mouse

Table of Contents

Related product manuals