EasyManua.ls Logo

Topcon TRC-NW8F - Maintenance; User and Manufacturer Maintenance

Topcon TRC-NW8F
104 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
88
MAINTENANCE
User maintenance items
Manufacturer maintenance items
Item Inspection time Contents
Inspection Before using • The instrument must operate correctly.
• The objective lens must be free of dirt or flaw,
etc.
Cleaning When the part is stained • Objective lens
• External cover, control panel, color LCD mon-
itor, etc.
Replacement As required • Illumination lamp
• Xenon lamp
•Fuse
Item Inspection time Contents
Cleaning each unit At least every 12 months • Cleaning the external parts
• Cleaning the optical system
• Cleaning the base unit
Operation check At least every 12 months • Operation of the instrument
• Operation of switches
Photography check At least every 12 months • Focus, flare, central ghost and dust
• Check the flash level.
• Check the illumination level.

Table of Contents