MAINTENANCE
User maintenance items
Manufacturer maintenance items
Item Inspection time Contents
Inspection Before using The instrument must operate correctly.
The objective lens must be free of dirt or flaw,
etc.
Cleaning When the part is stained Objective lens
External cover, control panel, color LCD mon-
itor, etc.
Replacement As required Illumination lamp
Xenon lamp
•Fuse
Item Inspection time Contents
Cleaning each unit At least every 12 months Cleaning the external parts
Cleaning the optical system
Cleaning the base unit
Operation check At least every 12 months Operation of the instrument
Operation of switches
Photography check At least every 12 months Focus, flare, central ghost and dust
Check the flash level.
Check the illumination level.

Table of Contents