91
MAINTENANCE
3 Loosen the three xenon set screws.
4 Hold the xenon PCB at the top and the bottom, slightly slide it to the right and pull it out
straight toward the operator side.
5 Insert the new xenon PCB, so the xenon lamp does not touch the surrounding metal com-
ponents, and upon reaching the stopper, slightly move it to the left and slide it into the
xenon set screws.
6 Securely fasten the three xenon set screws.
7 Attach the lamp house cover by matching the projection at the bottom part of the lamp
house cover with the groove of the body cover. Turn the screws with a coin, etc. and
securely fasten the lamp house cover.
To avoid electric shock, if the lamp house cover is left unfixed, an error is displayed on
the color LCD monitor and operations, including photography, cannot be done.
Xenon set screw
Xenon PCB