EasyManua.ls Logo

Toro Multi Pro 1200 - Troubleshooting; Sonic Boom Light; Ultra Sonic Boom Calibration

Toro Multi Pro 1200
326 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Multi Pro 1200/1250 Ultra Sonic Boom System (Rev. A)Page 9.1 - 17
Troubleshooting
For effective troubleshooting and repairs, there must be
a good understanding of the electrical circuits and com-
ponents used on the Ultra Sonic Boom System (see Ul-
tra Sonic Boom System Operation in this chapter).
NOTE: When troubleshooting an e lectrical problem on
your Ultra Sonic Boom System, refer to information r e-
garding the sonic boom light and diagnostic lamp in this
section. Also, use the Diagnostic Display (see Special
Tools in this chapter) to test Toro electronic controller
(TEC) inputs and outputs.
Sonic Boom Light
The sonic boom light is included in the sonic boom
switch on the spray control console (Fig. 7 or 8). This
light should be illuminated whenever the vehicle ignition
switch is ON and the sonic boom switch is in the auto-
matic position.
The sonic boom light flashing quickly indicates that the
Ultra Sonic Boom System is in the calibration mode.
This mode allows the spray booms t o be adjusted for the
desired boom height and continues for approximately
twenty (20) seconds. The sonic boom light will then flash
slowly for approximately two (2) minutes to finalize the
calibration settings.
NOTE: A sequence of switch movements is necessary
to engage the calibration mode. Refer to the Sonic
Boom Kit Installation Instructions for this sequence.
The sonic boom light flashes slowly when the sonic
boom switch is in the automatic position and a boom lift
switch is pressed to manually change the boom height.
The flashing light will return to being constantly ON and
automatic operation will be re-- engaged once the boom
switch is manually pressed to the lower position.
A slowly flashing sonic boom light may also indicate that
a s ystem fault has been encountered. In the event that
there is a fault in the Ultra Sonic Boom System (e.g.
there is no signal coming from a boom sensor), the af-
fected boom will raise briefly and then stop. The sonic
boom light will begin to flash slowly and the diagnostic
lamp on the console will also flash. If this occurs, refer
to Diagnostic Lamp, Diagnostic Display and Trouble-
shooting Chart in this section.
1. Spray control console 2. Sonic boom switch
Figure 7
1
2
MULTI PRO 1200
1. Spray control console 2. Sonic boom switch
Figure 8
1
2
MULTI PRO 1250
Ultra Sonic Boom Calibration
The sensor calibration process is critical to the correct
operation of the Ultra Sonic Boom System. The calibra-
tion process establishes the sensor target distance be-
tween the boom and the turf surface. Typically, this
distance is approximately twenty (20) inches. Steps
needed for proper calibration are identified in the Ultra
Sonic Boom Kit Installation Instructions.
While calibrating the Ultra Sonic Boom sensors, it is best
to perform the calibration process on turf. A shiny sur-
face (e.g. cement shop floor) can skew sensor signals.
Also, ensure the calibration area is free of buildings,
trees, underground plumbing and other machines that
could interfere with sensor signals.
Ultra Sonic
Boom System

Table of Contents

Other manuals for Toro Multi Pro 1200

Related product manuals