EasyManua.ls Logo

Transonic HT331-FT - Specifications

Default Icon
58 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AU-OPR-AureFloFT-EN,
Rev H
9
III. Specications
A. FlowTrace
®
Software
FlowTrace
®
Software is designed to work with a Transonic
®
safety-tested and approved Transonic
®
model
Monitor with the following minimum specications:
Windows XP,
Windows 7, or
Windows 8.1
1.6 GHz processor or higher
CD Rom drive 1 GB of RAM
Serial Port 20 GB Hard Drive
2 USB Ports Touch-panel display
Monitor display: 17” touch screen
B. HT300(-FT) Series Compatible Flowmeters
Single-channel (HT31x-FT & HT35x Series), dual-channel (HT32x-FT & HT36x Series) Flowmeters
Weight: Single Channel: 10 lb (4.5 kg); Dual Channel 11 lb (5 kg)
Dimensions all Models: 11 3/8” (28.9 cm) wide x 4” (10.2 cm) high x 13 1/2” (34.3 cm) deep
Attached two meter Flowprobe-to-Flowmeter extension cable per channel
Separate grounded medical grade power line-cord is supplied with the Flowmeter.
INPUT POWER
Universal Power Supply; 50-60 Hz; 100 -240 V ± 10% (automatic voltage adjustment)
FUSES
1.5A fast blo, mfg bussman: #GMA1.5, 250 VAC (HT31x-FT & HT32x-FT Series)
1A Slow Blo, mfg. Bel: # 5HT1-R 250VAC (HT35x & HT36x Series)
Both power entry conductors are protected by size 5 x 20 mm fuses.
ELECTRICAL ISOLATION
Cabinet is grounded; line to ground leakage current: less than 50 μA @100-120 V line; less than 100 μA @
220-240 V line.
All electronic components and cabling of the Flowprobe/Flowsensor extension cable, Flowprobe/
Flowsensor cable and Flowprobe/Flowsensor are fully isolated from the Flowmeter’s electrical circuitry to
meet IEC60601 “cardiac oating” specications.
Debrillator Protection: Class I “CF” IXPI device: Probes can be left attached to the patient during cardiac
debrillation for instantaneous report on the restoration of ow.
ENVIRONMENTAL
Operating temperature range: 1C to 40ºC
Operating, shipping and storage humidity: 0 to 90% RH non-condensing
Shipping and storage temperate range: -10ºC to 50ºC
RoHS compliant (HT35x & HT36x Series)
AUTOMATIC METER ADJUSTMENTS
Ultrasound frequency and insonication parameters
Flowprobe size and corresponding ow output ranges
Volume ow calibration of connected Flowprobe
Sampling rate optimized for local acoustic conditions

Table of Contents