AU-OPR-AureFloFT-EN,
Rev H
9
III. Specications
A. FlowTrace
®
Software
FlowTrace
®
Software is designed to work with a Transonic
®
safety-tested and approved Transonic
®
model
Monitor with the following minimum specications:
● Windows XP,
Windows 7, or
Windows 8.1
● 1.6 GHz processor or higher
● CD Rom drive ● 1 GB of RAM
● Serial Port ● 20 GB Hard Drive
● 2 USB Ports ● Touch-panel display
Monitor display: 17” touch screen
B. HT300(-FT) Series Compatible Flowmeters
● Single-channel (HT31x-FT & HT35x Series), dual-channel (HT32x-FT & HT36x Series) Flowmeters
● Weight: Single Channel: 10 lb (4.5 kg); Dual Channel 11 lb (5 kg)
● Dimensions all Models: 11 3/8” (28.9 cm) wide x 4” (10.2 cm) high x 13 1/2” (34.3 cm) deep
● Attached two meter Flowprobe-to-Flowmeter extension cable per channel
● Separate grounded medical grade power line-cord is supplied with the Flowmeter.
INPUT POWER
Universal Power Supply; 50-60 Hz; 100 -240 V ± 10% (automatic voltage adjustment)
FUSES
1.5A fast blo, mfg bussman: #GMA1.5, 250 VAC (HT31x-FT & HT32x-FT Series)
1A Slow Blo, mfg. Bel: # 5HT1-R 250VAC (HT35x & HT36x Series)
Both power entry conductors are protected by size 5 x 20 mm fuses.
ELECTRICAL ISOLATION
● Cabinet is grounded; line to ground leakage current: less than 50 μA @100-120 V line; less than 100 μA @
220-240 V line.
● All electronic components and cabling of the Flowprobe/Flowsensor extension cable, Flowprobe/
Flowsensor cable and Flowprobe/Flowsensor are fully isolated from the Flowmeter’s electrical circuitry to
meet IEC60601 “cardiac oating” specications.
● Debrillator Protection: Class I “CF” IXPI device: Probes can be left attached to the patient during cardiac
debrillation for instantaneous report on the restoration of ow.
ENVIRONMENTAL
● Operating temperature range: 15ºC to 40ºC
● Operating, shipping and storage humidity: 0 to 90% RH non-condensing
● Shipping and storage temperate range: -10ºC to 50ºC
● RoHS compliant (HT35x & HT36x Series)
AUTOMATIC METER ADJUSTMENTS
● Ultrasound frequency and insonication parameters
● Flowprobe size and corresponding ow output ranges
● Volume ow calibration of connected Flowprobe
● Sampling rate optimized for local acoustic conditions