AU-OPR-AureFloFT-EN,
Rev H
3
F. Perivascular Flowprobes & Clamp-on Tubing Flowsensors
Transonic
®
offers a wide range of intraoperative Perivascular Flowprobes and Extracorporeal Clamp-on
Tubing Flowsensors. Transonic
®
Flowsensors connect to a two-meter (~ 6.5 ft.) extension cable attached to
the front of the HT300 Series Compatible Flowmeter. Two or four ultrasonic transducers within the Sensor
body transmit a minimal level of ultrasound through a rectangular sensing window (Appendix E: Theory
of Operation, page 43). Volume ow of all non-aerated liquid passing through the sensing window is
measured.
1. PERIVASCULAR FLOWPROBES
Transonic
®
HQ-Series Intraoperative,
Perivascular Flowprobes work with HT300
Series Compatible Flowmeters to measure
instantaneous and average volume ow in
blood vessels or grafts, 0.5 mm to 36 mm in
diameter (Fig. 1.4). Perivascular Flowprobes
measure blood ow in exposed vessels:
a) at the beginning of a procedure, for accurate
assessment of volume ow.
b) at the conclusion of surgery prior to patient
closing, on newly implanted natural or
articial bypass grafts, on vessels supplying
restored, reconstructed or transplanted
organs and tissues.
2. CLAMP-ON FLOWSENSORS
Transonic
®
HQ-XL-Series Clamp-on Tubing Flowsensors apply ultrasound
energy through exible tubing to monitor instantaneous and average
volume ow of blood, cardioplegia or other solutions in (Fig. 1.5):
a) Extracorporeal procedures such as cardio-pulmonary bypass (CPB);
extracorporeal membrane oxygenation (ECMO); hemodialysis;
plasmapheresis; arteriovenous hemoltration (CAVH)
b) Extracorporeal perfusion, infusion or transfusion procedures (such
as allograft perfusion for pre-transplantation preservation; coronary
reperfusion or retroperfusion; continuous total nutrient infusion,
saline or dextrose infusion; blood transfusion)
c) Extracorporeal shunts (portal vein bypass shunts, or lower body/upper
body bypass shunts during liver transplants).
Fig. 1.4: An Array of Perivascular Flowprobes
Fig. 1.5: Clamp-on Flowsensor
Introduction