User's Manual
750 50 0042 03 60
By double clicking on a batch line (nominal value), it is
possible to view the results of all recorded measurements.
11.2.2 Statistical Values
The statistical values are displayed for each nominal value
of the batch. These are displayed as page header with
their tolerance (UTL & LTL).
To navigate between the nominal values, use the red
arrows.
The statistical values based on the results of each
nominal value are as follows:
Number of memorized results
N
Range
R
Xmax Xmin
Arithmetic mean value
X
N
Xi
N
i
=1
Standard deviation
σ
N
XXi
N
i
=
1
2
)(
Process capability
Cp
σ
6
LTLUTL