6–14
Press START/ENTER to begin taking the measurement. The display
will be:
SPECIAL CALIB.
Depth: x inches
Time:xxx secs.
After counting down to zero, the display will be:
Dens ct.=xxxx
Moist ct.=xxxx
To input sample
dens Press ENTER
Press START/ENTER to input the sample area wet density obtained
from the alternative test method. The display will be:
Input material
density -
x.x PCF
ENTER to ENABLE
Input the material density and press START/ENTER. The display will
return to READY.