APX 525/625 Generator Revision C Pre-Installation
©
2005 Del Medical Imaging Corporation
7/5/2005 Page 38
25. Loosen two stress relief clamp screws (1 Fig. 13).
26. Insert the end of the secondary cable (2) (P/N 126-0211G1) into the
stress relief clamp until the black insulation on the cable is within the
stress relief clamp as shown in Figure 13.
27. Tighten stress relief clamp screws.
28. Connect white and black wires to TB2 as shown below. It does not matter
which color wire goes in which position.
Figure 13. Secondary Cable Upper Connections
29. Reattach rear cover.
TB2
2
1