OAM CCM Results
RDI, LOC, XCON, UNEXP, and Alarm will display an Inactive or Active status.
Remote MPID: MEP ID of the remote MEP.
Remote MAC: MAC address of the remote MEP.
RDI: The CCM received contains the RDI flag set.
LOC: The MEP detects loss of connectivity.
XCON: Possible cross-connect, the CCM received could be from another MA.
UNEXP: Unexpected MEP ID or non matching CCM interval.
Alarm: A fault alarm is triggered if a defect is present for a time period of 10s. The fault alarm is cleared if a defect condition is
not present for a time period of 10s.
OAM - Service - LBM
OAM LBM Results
LBM Status
Pass: At least 1 Loopback response received
Fail: No Loopback responses received
To be Send: Outstanding number of LBM to be sent
Response Count