Cleaning the ion block assembly 5-45
20. Remove the extraction cone handle insulator from the extraction cone
handle.
21. Remove the extraction cone seal from the ion block.
Cleaning the ion block components
Required materials
Chemical-resistant, powder-free gloves.
Appropriately sized glass vessels in which to completely immerse
components when cleaning. Use only glassware not previously cleaned
with surfactants.
HPLC-grade (or better) methanol.
HPLC-grade (or better) water.
•Formic acid.
Ultrasonic bath.
Source of oil-free, inert gas (for example, nitrogen) for drying (air-drying
optional).
Wash bottle containing HPLC-grade (or better) 1:1 methanol/water.
Large beaker.
Extraction cone seal