2-6 Starting Up and Shutting Down the Mass Spectrometer
Shutting down the mass spectrometer
You can shut down the system by putting it in Standby mode, by fully
shutting it down, or by rebooting it.
Putting the mass spectrometer in Standby mode
Leave the mass spectrometer in Operate mode except in the following cases,
when you must put it in the Standby mode:
When performing routine maintenance
When changing the source
When leaving the mass spectrometer unused for a long period
To put the system in Standby mode
In the Tune window, click to put the mass spectrometer in
Standby mode.
Result: Doing so turns off the source voltages, gas flows, Intellistart
Fluidics system, and LC system.
Fully shutting down the mass spectrometer
To fully shut down the mass spectrometer
1. In the Tune window, click .
2. Click Vacuum > Vent.
3. Select Vent Instrument.
Result: A message confirms the vent command.
4. Click OK.
Result: When the turbomolecular pumps slow to half their normal
operating speed, the vent valves open, and the instrument automatically
vents.
5. Exit the MassLynx software.
6. Shut down the PC.