5-64 Maintenance Procedures
Installing the new capillary
Required materials
Chemical-resistant, powder-free gloves
Combined 2.5-mm Allen wrench and cone extraction tool
•10-mm wrench
Needle-nose pliers
•LC pump
HPLC-grade (or better) 1:1 acetonitrile/water
Capillary
Ferrule
•Seal
PTFE liner tubing
Conductive sleeve
Red PEEK tubing
Sharp knife or PEEK tubing cutter
Metal gasket for the probe tip
Safety goggles
To install the new capillary
Warning: The probe and source components can be contaminated
with biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning: The ESI probe tip is sharp. To avoid puncture wounds, handle
the probe with care.