EasyManua.ls Logo

Waters SYNAPT G2 - Maintenance Overview and Safety

Waters SYNAPT G2
236 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance schedule 5-3
Maintenance schedule
The following table lists periodic maintenance schedules that ensure optimum
instrument performance.
Maintenance schedule
Procedure Frequency For information...
Clean the instrument case. As required. See page 5-18.
Empty the nitrogen exhaust
trap bottle.
Check daily, empty as
required.
See page 5-19.
Replace the oil-free (scroll)
pump’s seals.
Annually. See Edwards
document XDS35i
Instruction Manual
A730-01-880.
Clean the source components. When sensitivity
decreases to
unacceptable levels.
See page 5-21.
Replace the ESI probe tip. When sensitivity
decreases to
unacceptable levels.
See page 5-55.
Replace the ESI probe
capillary.
When sensitivity
decreases to
unacceptable levels or
sample flow is
inconsistent.
See page 5-59.
Clean the APCI probe tip.
(Options using the APCI
probe only.)
When sensitivity
decreases to
unacceptable levels.
See page 5-67.
Replace the APCI probe
capillary.
When sensitivity
decreases to
unacceptable levels or
sample flow is
inconsistent.
See page 5-68.
Replace the LockSpray probe
capillary.
Annually. See page 5-74.

Table of Contents

Other manuals for Waters SYNAPT G2

Related product manuals