Rebooting the embedded PC 2-7
7. Switch off all the peripherals.
8. Switch off the vacuum, electronics, and the embedded PC and auxiliary
breakers located on the rear panel.
Rebooting the embedded PC
Reboot the embedded PC when either of these conditions applies:
The MassLynx software fails to initialize.
Immediately following a software upgrade.
To reboot the embedded PC
1. In the MassLynx software, close the Tune window.
2. On the instrument’s rear panel, switch off the EPC power switch, wait
5 seconds, and switch it back on.
3. Wait 4 minutes to allow full rebooting to take place.
4. Open the MassLynx software.
Caution: When rebooting, do not switch off power to the instrument, as
this will vent the instrument.