Replacing the ESI probe tip and gasket 5-57
Tip: If the probe tip is difficult to remove, use the 7-mm wrench in
conjunction with the 10-mm wrench.
3. Remove the metal gasket from the probe tip.
4. Dispose of the metal gasket in accordance with local environmental
regulations.
5. If the probe tip is damaged, dispose of it in accordance with local
environmental regulations.
Warning: The probe tip and metal gasket can be
contaminated with biohazardous and/or toxic materials.
Dispose of them according to local environmental
regulations.
Probe tip
7-mm wrench
10-mm wrench
Metal gasket